Boonton 4500B RF Peak Power Analyzer
Remote Operation
5-23
DISPlay:CHANnel:TOP: MIDdle
Description: Set or return the channel for the auto-measure or fixed parameter displayed in the Top row,
Middle column of the nine element parameter field above the trace box.
DISPlay:PARAmeter:MODE must be set to MEAS.
Syntax: DISPlay:CHANnel:TOP: MIDdle <asc>
Argument: <asc> = {CH1, CH2, TRIG1, TRIG2, CHM, MEM1, MEM2}
Valid Modes: Pulse Mode only. Note: Modulated mode does not display Status and Measurement parameters.
Statistical mode displays a fixed combination of status and auto-measure parameters.
DISPlay:CHANnel:TOP: RIGHt
Description: Set or return the channel for the auto-measure or fixed parameter displayed in the Top row, Right
column of the nine element parameter field above the trace box. DISPlay:PARAmeter:MODE
must be set to MEAS.
Syntax: DISPlay:CHANnel:TOP: RIGHt <asc>
Argument: <asc> = {CH1, CH2, TRIG1, TRIG2, CHM, MEM1, MEM2}
Valid Modes: Pulse Mode only. Note: Modulated mode does not display Status and Measurement parameters.
Statistical mode displays a fixed combination of status and auto-measure parameters.
DISPlay:PARAmeter:BOTtom:LEFT
Description: Set or return the auto-measure or fixed parameter displayed in the Bottom row, Left column of the
nine element parameter field above the trace box. The selected parameter will be for the currently
active channel. See DISPlay:ACTive command. DISPlay:PARAmeter:MODE must be set to
MEAS.
Syntax: DISPlay:PARAmeter:BOTtom:LEFT <asc>
Argument: See Display Parameter Table below
Valid Modes: Pulse Mode only. Note: Modulated mode does not display Status and Measurement parameters.
Statistical mode displays a fixed combination of status and auto-measure parameters.
DISPlay:PARAmeter:BOTtom:MIDdle
Description: Set or return the auto-measure or fixed parameter displayed in the Bottom row, Middle column of
the nine element parameter field above the trace box. The selected parameter will be for the
currently active channel. See DISPlay:ACTive command. DISPlay:PARAmeter:MODE must be
set to MEAS.
Syntax: DISPlay:PARAmeter:BOTtom:MIDdle <asc>
Argument: See Display Parameter Table below
Valid Modes: Pulse Mode only. Note: Modulated mode does not display Status and Measurement parameters.
Statistical mode displays a fixed combination of status and auto-measure parameters.