Boonton 4500B RF Peak Power Analyzer
Operation
Table 4-7C. Ext Trig # > Calibration > Submenu
Menu Item Function
(Type) Selections (SCPI cmd)
AutoCal Pulse, Mod & Stat Initiates the automatic trigger channel
(Action) START calibration routine
Disconnect all signal sources from the
External Trigger input to be calibrated. Press
the Autocal START menu key to initiate a
zero and amplitude calibration.
Trigger channel zero and calibration
information is saved in a system file and
restored during initialization.
CALibration[3|4]:AUTO
Zeroing Pulse, Mod & Stat Initiates the automatic trigger channel
(Action) START zero adjustment
Disconnect all signal sources from the
External Trigger input to be zeroed. Press
the Zeroing START menu key to initiate
a zero.
CALibration[1|2]:ZERO