Boonton 4500B RF Peak Power Analyzer
Operation
4-56
Table 4-12. Mark > Extensions > Submenu (continued)
Menu Item
(Type) Selections Function
the two markers. If Mark > Extensions > Delta Marker 
and the units are changed to logarithmic, the selection is automatically
switched to the previous selection of Ratio or Avg. If the units are
-selected.
The MK1-MK2 and MK2-MK1 selections for marker math are valid
for the above functions. The MIN-MAX, MAX-MIN and Pk/Avg
marker math functions are available only when Mark > Extensions >
Mk Group is set to Both. In addition, Pk/Avg requires that Mark >
Extensions > Delta Marker is set to 
Stat In the Statistical mode 
Ratio units. Only the MK1-MK2 selection for marker math is valid for this
function.
See Marker Math SCPI commands.
Marker Math Pulse & Mod Selects the type and order of measurement markers
(Mult. Choice)
Marker 1 and Marker 2 measure the signal level at the set point on the
waveform.
MK2 - MK1 Ratio = MK2/MK1, Delta = MK2 MK1
MK1 - MK2 Ratio = MK1/MK2, Delta = MK1 MK2
MIN and MAX measure the minimum and maximum instantaneous
signal level between the markers.
MAX MIN Ratio = MAX/MIN, Delta = MAX MIN
MIN MAX Ratio = MIN/MAX, Delta = MIN MAX
MINF and MAXF measure the minimum and maximum filtered signal
level between the markers.
MAXF MINF Ratio = MAXF/MINF, Delta = MAXF MINF
MIN F MAXF Ratio = MINF/MAXF, Delta = MINF MAXF
Pk/Avg Ratio = Pk/Avg
LAvg measures the long term signal average between the markers.
PkHld measures and holds the peak signal value. See the
Chan # > Exten > Peak Hold function for more information.
Stat Selects the type and order of statistical measurement markers
Marker 1 and Marker 2 measure the signal level at the set points on the
distribution.
MK1 - MK2 Ratio = MK1/MK2 (only)
Przeglądanie stron 127