Step 9: The Sensor Calfactor Uncertainty needs to be interpolated from the uncertainty
values given in Table 3-2 (Peak Power Sensor Calibration Factor Uncertainty). At 1 GHz, the
sensor’s calfactor uncertainty is 1.7 %, and at 0.5 GHz it is 1.6 %. Note, however, that
we are performing our AutoCal at a frequency of 1 GHz, which is very close to the
measurement frequency. This means that the calfactor uncertainty cancels to zero at 1 GHz.
We’ll use linear interpolation between 0.5 GHz and 1 GHz to estimate a value. 900 MHz
is only 20% (one fifth) of the way from 1GHz down to 500MHz, so the uncertainty figure at
0.5 GHz can be scaled by one fifth.
U
CalFacto
r
= [ ( F - F
1
) * (( CF
2
- CF
1
) / ( F
2
- F
1
)) ] + CF
1
where; F = 0.9
F
1
= 0.5 CF
1
= 1.6
F
2
= 1.0 CF
2
= 0.0
= [ ( 00.9 - 00.5 ) * (( 0.0 - 1.6 ) / ( 1.0 - 0.5 )) ] + 1.6
= [ ( 0.4 ) * (( -1.6 ) / ( 0.5 )) ] + 1.6
= [ ( 0.4 ) * ( -1.6 ) ] + 1.6
= 0.32 %
Step 10: Now that each of the individual uncertainty terms has been determined, we can
create an uncertainty budget and calculate the combined standard uncertainty (Uc) .
Source of Symbol
Value
Probabilty
Divisor Ustd
Uncertainty
(+/- %)
Distribution
(+/- %)
Instrument
I 0.2 normal
2 0.10
Calibrator
Level
R 3.11 rectangular
( 3 )
0.5
1.80
Mismatch
M
C
1.27 U-shaped
( 2 )
0.5
0.90
Source
Mismatch
M
S
0.80 U-shaped
( 2 )
0.5
0.57
Sensor
Shaping
S 2.00 rectangular
( 3 )
0.5
1.15
Temp. Drift
T 1.69 rectangular
( 3 )
0.5
0.98
Noise
N 0.03 normal
2 0.02
Cal Factor
K 0.32 normal
2 0.16
Combined Standard
Uc normal
2.58
Uncertainty
Expanded
U normal
5.17
Uncertainty
(k=2)
From this example, different uncertainty terms dominate. Since the measurement is close t
o
the calibration frequency, and matching is rather good, the shaping and level errors are the
largest. The Expanded Uncertainty of 5.17 % translates to an uncertainty of about 0.22 dB
in the reading.