INSTRUCTION MANUAL 4240 SERIES RF POWER METER REV DATE 01/19/2011 MANUAL P/N 98406700A 06799A CD P/N 984

Strona 2

Boonton 4240 Series RF Power Meter This page intentionally left blank. Contents x

Strona 3 - SAFETY SUMMARY

Boonton 4240 Series RF Power Meter 5-1. SCPI COMMAND SUMMARY (continued) SENSe[1|2]:CORRection:CA<numeric value> = -3.00 to 3.00 et/return dut

Strona 4

Boonton 4240 Series RF Power Meter Table 5-1. SCPI COMMAND SUMMARY (continued) STATus:QUEStionable:CALibration:EVENt? Ret urn value

Strona 5

Boonton 4240 Series RF Power Meter 5.5.16 4230 Emulation GPIB Commands Table 5-2. 4230 EMULATION GPIB COMMANDS Code Description Comments CP

Strona 6

Boonton 4240 Series RF Power Meter Table 5-2. 4230 EMULATION GPIB COMMANDS (continued) Code Description Comments Channels ngle channel al, fr

Strona 7

Boonton 4240 Series RF Power Meter 5.5.17 HP 437B Em ulation GPIB Commands Table 5-3. HP 437B EMULATION GPIB COMMANDS UPPORTED RG # 100) NOT

Strona 8

Boonton 4240 Series RF Power Meter Table 5-3. HP 437B EMULATION GPIB COMMANDS Code D s KZ ction function NOT SUPPORTED llator off n t

Strona 9

Boonton 4240 Series RF Power Meter Table 5-3. HP 437B EMULATION GPIB COMMANDS Code Description Comments SM Status message NOT SUPPOR

Strona 10 - Contents

Boonton 4240 Series RF Power Meter 5.5.1 ands 8 HP 438A Emulation GPIB CommTable 5-4. HP 438A EMULATION GPIB COMMANDS Comments D Measure A-

Strona 11 - 1. General Information

Boonton 4240 Series RF Power Meter LP1 Set learn mode #1 NOT SUPPORTED Table 5-4. HP 438A EMULATION GPIB COMMANDS ode Description Comm

Strona 12

Boonton 4240 Series RF Power Meter 6. Application Notes in this section include pulse measurement fundamentals, automatic measurement rinciples, and a

Strona 13 - General Information

Boonton 4240 Series RF Power Meter 1. General Information This instruction manual provides you with the information you need to install, operate and

Strona 14

Boonton 4240 Series RF Power Meter Figure 6-1. Pulsed RF Signal Figure 6-2. Distorted Pulse Signal Application Note

Strona 15

Boonton 4240 Series RF Power Meter 6.1.2 Diode Detection Wideband diode detectors are the dominant power sensing device used to measure pulsed RF

Strona 16 - Sensor dependant )

Boonton 4240 Series RF Power Meter Frequenccapacicannottable f Misma nding on the mismatminim Signalcalibrat nal and second harmonic combininsigni egi

Strona 17

Boonton 4240 Series RF Power Meter 6.2 M The 424 Standard year calibratio chapter assumes that the power meter is being maintained correctly and is

Strona 18

Boonton 4240 Series RF Power Meter Tual his reason, the uncertainties are more commonly combined using the RSS method. RSS is an abbreviation for “ro

Strona 19

Boonton 4240 Series RF Power Meter The value to use for calibration level uncertainty depends upon the sensor calibration technique used. If AutoCal

Strona 20

Boonton 4240 Series RF Power Meter Sensor Temperature Coefficient. This term is the error which occurs when the sensor’s temperature has changed sign

Strona 21 - 2. Installation

Boonton 4240 Series RF Power Meter 6.2.3 Sample Uncertainty Calculations. The following example shows calculations for a CW power sensors. The fig

Strona 22 - Installation

Boonton 4240 Series RF Power Meter Step 6: The Sensor Temperature Drift Error depends on how far the temperature has drifted from the sensor calibrati

Strona 23 - 2.4 Preliminary Check

Boonton 4240 Series RF Power Meter Step 1 certainty terms has been determined, we can combine them to calculate the worst-case a U (±%) K (U×K)2

Strona 24

Boonton 4240 Series RF Power Meter 1.2 Description The Model 4240 is a digital signal processor based, single or dual channel, solid state RF power

Strona 25 - 3. Getting Started

Boonton 4240 Series RF Power Meter This page intentionally left blank. Application Notes 6-12

Strona 26 - Getting Started

Boonton 4240 Series RF Power Meter 7. Maintenance Th .1 Safety

Strona 27

Boonton 4240 Series RF Power Meter 7.5 Firmware Upgrade Instructions equirements The Firmware Update file is a text file in S-record format. Any uti

Strona 28

Boonton 4240 Series RF Power Meter 7. During time the 4240 will display "FIRMWARE DOWNLOAD". Thapprox e 45 seconds to completely transfer.

Strona 29

Boonton 4240 Series RF Power Meter This page intentionally left blank. Maintenance 7-4

Strona 30 - 3.3 Operation

Boonton 4240 Series RF Power Meter 8. ges Appendix A SCPI Error Messa .1 SCPI Erro8 r Messages -101 “SubCmd not found” -113 "Undefined head

Strona 31

Boonton 4240 Series RF Power Meter NO. MESSAGE DESCRIPTION sor Disconnected." -249 "FPGA download err" 63 “MFS Init” 64 “Flash i

Strona 32

Boonton 4240 Series RF Power Meter Appendix B 9-1 9. Appendix B Warranty & Repair Repair Policy 4240 Series Instrument. If the Boonton 4240 Ser

Strona 33

Boonton 4240 Series RF Power Meter Appendix B 9-2 THE FOREGOING WARRANTIES ARE IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUD-ING, BUT

Strona 34

Boonton 4240 Series RF Power Meter • ual Independent Channels. When equipped with the optional second measurement channel, the • • optimized throu

Strona 35

Boonton 4240 Series RF Power Meter • • ne key press operations. To provide for ease of use operation of the instrument functions that mply by pres

Strona 36

Boonton 4240 Series RF Power Meter 1.4 Accessories 40 ac data sensor adapter for each channel installed the instrument. One or more Boonton 51000

Strona 37

Boonton 4240 Series RF Power Meter Option designations are appended to the instrument’s base model number. For example, Model 4242-01-nel instrum ca

Strona 38

Boonton 4240 Series RF Power Meter Table 1-2 4240 Series Performance Specifications (continued) pecifications aNCERTAINTIES easurement Accuracy Sum

Strona 39

Boonton 4240 Series RF Power Meter Table 1-2 4240 Series Performance Specifications (continued) (Specifications are subject to change without notic

Strona 40

Boonton 4240 Series RF Power Meter Table 1-2 4240 Series Performance Specifications (continued) (Specifications are subject to change without notic

Strona 41

Boonton 4240 Series RF Power Meter INSTRUCTION MANUAL, 4240 SERIES RF POWER METER Revision date 01/19/2011 © Cop

Strona 42

Boonton 4240 Series RF Power Meter This page intentionally left blank. General Information 1-10

Strona 43

Boonton 4240 Series RF Power Meter 2. Installation This section contains unpacking and repacking instructions, power requirements, connection descrip

Strona 44

Boonton 4240 Series RF Power Meter Table 2-1 4240 Series Packing List INSTRUMENT (See also Table 1-1) 4240 Series RF Power Meter Line Cord Boon

Strona 45

Boonton 4240 Series RF Power Meter Recorder If a recorder is to be used to record measurement data, connect the recorder to the recorder BEC conne

Strona 46

Boonton 4240 Series RF Power Meter 6. Press the <MENU> key and select DIAGNOSTICS with the down arrow key. Press <ENTER>.

Strona 47 - 4. Operation

Boonton 4240 Series RF Power Meter 3. Getting Started This chapter will introduce the user to the 4240 Series. The chapter will identify objects on

Strona 48 - Operation

Boonton 4240 Series RF Power Meter Figure 3-1. Standard 4240 Series RF Power Meter - Front Panel Table 3-1 Operating Controls, Indicators and C

Strona 49

Boonton 4240 Series RF Power Meter Table 3-1 Operating Controls, Indicators and Connections (continued) Reference # Front Rear Nomenclature Fu

Strona 50

Boonton 4240 Series RF Power Meter Table 3-1 Operating Controls, Indicators and Connections (continued) Reference # Front Rear Nomenclature Fu

Strona 51 - 4.5 Noise

Boonton 4240 Series RF Power Meter Figure 3-2. 4240 Series - Rear Panel (Shown without optional rear panel connectors not installed) Getting Star

Strona 52

Boonton 4240 Series RF Power Meter SAFETY SUMMARY The following general safety precautions must be observed during all phases of operation and main

Strona 53

Boonton 4240 Series RF Power Meter 3.3 Operation The Model 4 c e r e o i e c h o l 240 an b configu ed for op rati n v a th six swit hes on

Strona 54

Boonton 4240 Series RF Power Meter DUAL CHANNEL ACM M M M M L U PK∆ ± D D D D D D U U U ( B A R G R A P H )

Strona 55

Boonton 4240 Series RF Power Meter 3-5. Mod nd Set ast Menu Operatio ey stokes to perform a function repeat d , the control program can ing to t .

Strona 56 - .6 Measurement Time

Boonton 4240 Series RF Power Meter 3.3.1 Menu Key. he instrument's, CHA P, REPORT and DIAGNOSTIC functions are accessed when the sor can be posit

Strona 57 - 4.7 High Frequ

Boonton 4240 Series RF Power Meter Table 3-2 CHANNEL MENU Functions Function De Parameters Defaults RTN e n e tr e o r u e . NITS Units used for

Strona 58 - 4.8 Waveform Sensitivity

Boonton 4240 Series RF Power Meter R T N L E L 6 V E > - 0

Strona 59

Boonton 4240 Series RF Power Meter Table 3-4 SETUP MENU Functions n u Function De rs Defaults SAVE Saves the current instrument 1- 0 1 confi

Strona 60 - 4.10 Bar Graph Operation

Boonton 4240 Series RF Power Meter Table 3-5 IEEE MENU Functions (continued) OS LSTN End of string indicator for LF, CR, CRLF, NONE LF received m

Strona 61 - 5. Remote Operation

Boonton 4240 Series RF Power Meter Table 3-6 RS232 MENU Functions (continued) Function Description rs e u PARITY bit mode in a m ssage. ODD, NO

Strona 62 - 5.3 SCPI Language

Boonton 4240 Series RF Power Meter Table 3-7 DIAGNOSTICS MENU Functions (continued) Function Description Parameters Defaults SELF TEST Instructs

Strona 63

Boonton 4240 Series RF Power Meter SAFETY SYMBOLS This safety requirement symbol (located on the rear panel) has been adopted by the International El

Strona 64

Boonton 4240 Series RF Power Meter Figure 3-15. Access Code R T N M &

Strona 65

Boonton 4240 Series RF Power Meter Function Description Parameters Defaults : Factor 5000 [0 to 6] : [0 to 9999] the n/a n/a calibratio

Strona 66 - 5.5 SCPI Command Reference

Boonton 4240 Series RF Power Meter Table 3-9 FREQ Menu Functions unction Description Parameters Defaults F FREQ Frequency 0.01 to 100.00 GHz

Strona 67

Boonton 4240 Series RF Power Meter 3.3.3 FREQ Key. he freTinquency of the signal being measured must be entered in order to use the stored high frequ

Strona 68

Boonton 4240 Series RF Power Meter he user may also perform a 0 dBm calibration by one key stT roke of the Zero/Cal key. Simply connect the sensor to

Strona 69

Boonton 4240 Series RF Power Meter Table 3-10 Reference Level Menu Functions unction Description Parameters Defaults Bm Reference level value in

Strona 70

Boonton 4240 Series RF Power Meter This page intentionally left blank. Getting Started 3-22

Strona 71

Boonton 4240 Series RF Power Meter 4. Operation This section provides detailed background information on various aspects of operation of the Model 4

Strona 72

Boonton 4240 Series RF Power Meter these figures show the decay of measured power after a large signal has been applied. The typical error that can b

Strona 73

Boonton 4240 Series RF Power Meter Figure 4-2 Diode Sensor Decay Operation 4-3

Strona 74

Boonton 4240 Series RF Power Meter 1. General Information...

Strona 75

Boonton 4240 Series RF Power Meter 4.3 Dynamic Range The hold range mode is useful when he hold range mode is active when a specific instrument e

Strona 76

Boonton 4240 Series RF Power Meter 4.4 Filtering The Model 4240 employs digital filtering (averaging of measurements) to reduce the noise floor of

Strona 77

Boonton 4240 Series RF Power Meter Operation 4-6

Strona 78

Boonton 4240 Series RF Power Meter Figure 4-5. Noise Reduction Figure 4-6. Typical Error Band Specifications Figure 4-7. Probability of Falling wi

Strona 79

Boonton 4240 Series RF Power Meter Figure 4-8. Confidence Curves, 51013 Sensor with 2.8 Second Filter Operation 4-8

Strona 80

Boonton 4240 Series RF Power Meter Figure 4-9. Confidence Curves, 51013 Sensor with 10 Second Filter Operation 4-9

Strona 81

Boonton 4240 Series RF Power Meter Noise Error Examples. Figures 4-8 and 4-9 show the computed error for the 51013 diode sensor at different power le

Strona 82

Boonton 4240 Series RF Power Meter ontinuous Response. Regardless of the overhead time or the digital filter length, the Model 4240 will output readi

Strona 83

Boonton 4240 Series RF Power Meter Figure 4-11. Mismatch Uncertainties Chart 4.8 Waveform Sensitivity Thermal sensors are insensitive to the wav

Strona 84

Boonton 4240 Series RF Power Meter Diode 20 dBm (-10 dBm and 0 dBm for attenuated models 51015 and 1033). This response characteristic is obtained b

Strona 85

Boonton 4240 Series RF Power Meter 4.3 Dynamic Range...

Strona 86

Boonton 4240 Series RF Power Meter nly the display is affected by the duty cycle calculation. The mOperation 4-14 easurement process is subject to t

Strona 87

Boonton 4240 Series RF Power Meter 5. Remote Operation 5.1 GPIB Configuration The 4240 GPIB interface is configured using the Menu > SETUP >

Strona 88

Boonton 4240 Series RF Power Meter Additionally, the ASCII "DC2" character (hexadecimal 12/CTRL R) will cause the instrument to immediately

Strona 89

Boonton 4240 Series RF Power Meter suffix is not present, the default channel is assumed. strument’s “Channel 1” measurement path, while SENSFor exam

Strona 90

Boonton 4240 Series RF Power Meter Examples: :CALCulate:STATe ON Turn on measurement channel 1 (default channel number) ON Turn on measurement chan

Strona 91

Boonton 4240 Series RF Power Meter 5.4 Basic Measurement Information e easiest asequence which includes a MEAS1:POWERMEAS1:VOLTAGE? To return th

Strona 92

Boonton 4240 Series RF Power Meter 5.5 SCPI Command Reference te commands accepted by the 4240 Series. The list is grouped by SCPI subsystem r IEEE

Strona 93

Boonton 4240 Series RF Power Meter *ESR? Description: Return the current value of the Standard Event Status Register, then clear the register. Th

Strona 94

Boonton 4240 Series RF Power Meter *OPT? Description: Return the status of Channel 1 and Channel 2 followed by a list of installed options. Syntax

Strona 95

Boonton 4240 Series RF Power Meter *STB? Description: Return the current value of the Status Byte register. This register has bits assigned to a

Strona 96

Boonton 4240 Series RF Power Meter CALCulate:LIMit:LOWer[:POWer] ... 5-10 CALCulate:LIMi

Strona 97

Boonton 4240 Series RF Power Meter 5.5.2 CALCulate Subsystem he CALCulate group of the command subsystem is used to configure post acquisition data p

Strona 98

Boonton 4240 Series RF Power Meter CALCulate:LIMit:UPPer[:POWer] Description: Set or return the upper limit power level for the selected channel. Thi

Strona 99

Boonton 4240 Series RF Power Meter CALCulate:MATH:ARGA Description: Set or return the first argument to be used for channel math operations.

Strona 100 - Remote Operation

Boonton 4240 Series RF Power Meter CALCulate:REFerence:COLLect Description: For the selected channel, make the current power the reference level

Strona 101

Boonton 4240 Series RF Power Meter 5.5.3 CALibration Subsystem ommands is used to control automatic zero offset and linearity adjustments to the RF p

Strona 102

Boonton 4240 Series RF Power Meter 5.5.4 DISPlay Subsystem Syntax: DISPlay:ACTive <numeric value> Argument: <numeric value> = 1 to 2

Strona 103

Boonton 4240 Series RF Power Meter DISPlay:LABel:TEXTA Description: Displays a text message of up to 20 characters in the first label field when Di

Strona 104

Boonton 4240 Series RF Power Meter 5.5.5 FETCh Queries The FETCh? group of queries is used to return specific measurement data from a measurement cyc

Strona 105

Boonton 4240 Series RF Power Meter 5.5.6 INITiate and ABORt Commands tion in progress and prepares e instrument for an INITiate command. BORt on:

Strona 106

Boonton 4240 Series RF Power Meter 5.5.7 MEASure Queries used to acquire data using a set of high level instructions. They are structured to allow

Strona 107

Boonton 4240 Series RF Power Meter OUTPut:LEVel[:POWer] ... 5-21 OUTPut

Strona 108

Boonton 4240 Series RF Power Meter 5.5.8 MEMory Subsystem is used to save and recall instrument operating configurations, and to edit and review use

Strona 109 - 6. Application Notes

Boonton 4240 Series RF Power Meter 5.5.9 OUTPut Subsystem The OUTPut group of commands is used to control various outputs of the 4240. These output

Strona 110 - Application Notes

Boonton 4240 Series RF Power Meter OUTPut:RECorder:SOURce Description: Set or return the source channel for the Recorder Output. Syntax: OUTPut:R

Strona 111

Boonton 4240 Series RF Power Meter 5.5.10 READ Queries he purp AData. READ s theis equivalent to ABORtinready. READ? generally every time it is iss

Strona 112

Boonton 4240 Series RF Power Meter 5.5.11 SENSe Subsystem The purpose of the SENSe command subsystem is to directly configure device specific setting

Strona 113

Boonton 4240 Series RF Power Meter SENSe:FILTer:STATe Description: Set or return the current setting of the integration filter on the selected cha

Strona 114

Boonton 4240 Series RF Power Meter 5.5.12 STATus Commands ables you to control the SCPI defined status reporting structures. The user may amine the

Strona 115

Boonton 4240 Series RF Power Meter STATus:DEVice:EVENt? Description: Returns the current contents of the Device Event register then resets the reg

Strona 116

Boonton 4240 Series RF Power Meter STATus :CO Description: real time, and can change es in the instrument’s operating condition. :OPERation NDition

Strona 117

Boonton 4240 Series RF Power Meter STATus:OPERation:NTRansition alue of the negative transition filter bitmask for the Operation status group. the n

Strona 118

Boonton 4240 Series RF Power Meter SYSTem:COMMunicate:SERial:PARity... 5-34 SYSTem:COMMunicate:

Strona 119

Boonton 4240 Series RF Power Meter STATus:QUEStionable:CONDition? f the Questionable Condition register. The following table shows the Bit Va Descr

Strona 120

Boonton 4240 Series RF Power Meter STATus:QUEStionable:NTRansition Description: Set or return the value of the negative transition filter bit mask

Strona 121 - 7. Maintenance

Boonton 4240 Series RF Power Meter STATus:QUEStionable:CALibration:ENABle Description: Sets or returns the Questionable Calibration Enable register

Strona 122 - Maintenance

Boonton 4240 Series RF Power Meter 5.5.13 SYSTem Subsystem T Tem group o mmands is used to control system-level functions not directly related to ins

Strona 123

Boonton 4240 Series RF Power Meter SYSTem:COMMunicate:GPIB:TALK of-string terminator . UD UD <numeric value> 00, 1200, 2400, 4800, 9600, 19200,

Strona 124

Boonton 4240 Series RF Power Meter SYSTem:ERRor[:NEXT]? d error code number followed by a quoted ASCII text string describing the error. Note that er

Strona 125 - Appendix A SCPI Error Messa

Boonton 4240 Series RF Power Meter .5.14 INSTrument:VERSion Commands p of commands is used to q ode follows a “YYYYMMDD” format. The FPGA revisio s

Strona 126 - Appendix A

Boonton 4240 Series RF Power Meter .5.15 SCPI Command Summary Table 5-1. SCPI 5 COMMAND SUMMARY ESE Set/get Standard EvenBORt to idle iate] L?

Strona 127 - HTwww.boonton.comTH

Boonton 4240 Series RF Power Meter Ta d) ble 5-1. SCPI COMMAND SUMMARY (continue CALCulate:MATH:ARGB Set or return the second argument to be used fo

Strona 128 - Appendix B

Boonton 4240 Series RF Power Meter Table 5-1. SCPI COMMAND SUMMARY (continued) DISPlay:Label:TEXTC Displays a text message in the third label field.