13.2 Performance Check
To maintain maximum performance, it is highly recommended to conduct a basic
test of the PIM31 on a daily basis.
The following components are required for the performance test:
Low-IM Termination Load (Boonton 50W Low PIM Load)
Low PIM Cable
Install the test set-up as shown:
13.2.1 Testing Analyzer operation status
Connect the Low PIM Load to a PIM 31 test system with torque wrench
Set PIM 31 to Analyzer Mode
Set the frequencies of carrier signals 1 and 2.
Set the power levels to 43 dBm (20W) with ALC: ON