12.2 Specifications (Data Sheet)
12.2.1 Transmitter Specification
Carrier Power Adjustable Level +20 to + 44 dBm
Carrier Power Resolution / Accuracy 0.25 dB / ±0.35 dB
Frequency Range see version table
Frequency Increment 200 kHz
Frequency Accuracy (typical) 2 ppm
Frequency Tuning Lock Time (typical) 1 ms
Reverse Power Protection 43 dBm / 5 sec
(Include ON/OFF Function)
12.2.2 Receiver Specifications
Reverse IM -132 dBm / -175 dBc
Accuracy ±0.35 dB
Average Noise Floor -138 dBm (1.2 k filter)
Dynamic Range (typical) 96 dB
(reference -90 dBm)
Measurement Interval 100 to 350 ms
Effective IF Bandwidth 300, 600, 1.2k, 2.4k,
(Operator selectable) 5k, 10k, 12k, 15k, 25k
50kHz
Operational Input Power - 45 dBm RMS
Input Power without damage - 10 dBm max
Warm-Up Time for specified accuracy 5 minutes
12.2.3 System Specifications
RF Connector DIN 7-16
(3) USB ports
VGA output
REF Out 10MHz
User Interface Display 7” TFT with touch screen display
IM Measurement Alarms Audio / visual
Operating Voltage 90 to 264 V (50-60 Hz)
Power requirements 750 VA max