AC Supply: 90-264V, 750W / 750VA
use 90 deg connector only.
Fuse: 4A / 230V or 8A / 110V.
PIM31 contains protection circuitry to control
RF power during operation and during power
on/off cycles.
After switching Main Power Switch on, wait 2
seconds before pushing Front Power Button.
Fans may briefly run after power off.
Note: Do not use Main Power Switch while
system is in operation mode as vital files
may be corrupted. Always use the Front
Power Button to power the system down.
10 MHz reference output to synchronize
external equipment.
Always ensure sufficient air flow.
Always ensure sufficient air flow.
Getting Started
(1) Powering Up:
PIM 31 test systems power up in a similar fashion to a desktop PC. However,
there is one exception: Significant RF power has to be controlled and managed.
For this reason, PIM 31 Series instruments contain a system to protect the
hardware and prevent unusual on / off cycles (such as 3 cycles per second).
When operated normally, the user will not be aware of these protective
measures.
• Briefly press the start button on the front panel.