Boonton 4540 Series RF Power Meter
Remote Operation
OUTPut:MIO:STATus:SETTing
Description: Set/return Multi IO status out setting. Below is a description of each argument:
Argument Description
ALMACT: Sets the Multi IO output to +5 VDC when an alarm is currently active.
Multi IO output will be 0 VDC if no alarms are active.
ALMLATCH: Sets the Multi IO output to +5 VDC when an alarm limit has been
exceeded. Multi IO output will be 0 VDC if no alarm limits have been
exceeded.
MEASRDY: Multi IO output is 5 VDC when a measurement is ready to be read.
CAL: Multi IO output is 5 VDC when a calibration is complete.
Syntax: OUTPut:MIO:STATus:SETTing <character data>
Argument: <character data> = { ALMACT, ALMLATCH, MEASRDY, CAL }
Valid Modes: Any
OUTPut:MIO:STATus:SOURce
Description: Set/return Multi IO source for status out (alarm). Single channel units (Model 4541) only accept
―CH1‖.
Syntax: OUTPut:MIO:STATus:SOURce <character data>
Argument: <character data> = { CH1, CH2, BOTH }
Valid Modes: Any