Boonton 4540 Series RF Power Meter
Remote Operation
5-30
FETCh:TEMPerature:AVERage?
Description: Return the averaged internal temperature of the specified peak power sensor. Note: If a non-
temperature compensated peak sensor is in use, the 4540 displays a warning message if the
temperature has drifted more than 4C from the autocal temperature.
Syntax: FETCh[1|2]:TEMPerature:AVERage?
Returns: CC, Sensor[1|2] Average Temp in degrees C
Where CC is the measurement condition code.
Valid Modes: Peak Sensors only
FETCh:TEMPerature:CURRent?
Description: Return the current internal temperature of the specified peak power sensor. Note: If a non-
temperature compensated peak sensor is in use, the 4540 displays a warning message if the
temperature has drifted more than 4C from the autocal temperature.
Syntax: FETCh[1|2]:TEMPerature:CURRent?
Returns: CC, Sensor[1|2] Temp in degrees C
Where CCn is the measurement condition code.
Valid Modes: Peak Sensors only
FETCh:TEMPerature:INTernal?
Description: Return the current internal temperature of the instrument in degrees Celsius.
Syntax: FETCh:TEMPerature:INTernal?
Returns: CC, Instrument Temp in degrees C
Where CCn is the measurement condition code.
Valid Modes: all